MULTI-PARAMETER PATIENT MONITOR MODEL : MPPM1001

DISPLAY
DISPLAYED PARAMETERS
ECG
Protected against defib. & electro surgery potentials
SpO2
PULSE RATE
RR
TEMP
NIBP
POWER REQUIREMENTS
ENVIRONMENT
The transport and storage environment should comply with the following conditions:
OPTIONAL ACCESSORIES
Thermal Printer
IBP Modules
SPO2 Sensors
NIBP Cuffes
2nd Temperature Channel / Probe
Type: |
640 X 480 pixel color TFT |
Screen Size: |
10.4 inch diagonal |
DISPLAYED PARAMETERS
Time: |
Battery-backed quartz crystal clock |
Alarms: |
Hi & lo limits selectable on patient para. |
ECG: |
ECG Waveform Scale, displayed lead |
Heart Rate: |
Derived from ECG |
NIBP: |
Pressure (systolic, mean and diastolic) |
Pulse Ox: |
Pulse Rate, Pulse waveform, & % sat |
RR Apnea |
Respiration rate derived from ECG. |
Trends: |
HR, RR, NIBP, Temp.& SpO2 |
Temperatures: |
One channel (2nd channel optional) |
Trace Freeze: |
Traces A or B |
ECG
Protected against defib. & electro surgery potentials
Standard Lead: |
I, II, III, AVL, AVR, AVF, V1, V2, V3, V4, 5,V6 |
Display |
5, 7, 10,15, 20, 25mm/mV |
Gain Scales |
|
Sweep Speed |
12mm/s, 25mm/s, 50mm/s |
Input Resis. |
> 5M Ohm |
Freq. Response |
0.05Hz-100Hz (3dB) |
C M R R |
>80dB |
Electode Offset |
Maximum ±0.3V |
Potential Baseline Recovery |
<5s after 5KV defibrillation |
Heart Rate Range |
15 to 300bpm |
Resolution |
1bpm |
Accuracy |
<100bpm ±1% =100bpm ±2% |
Alarm |
Heart rate hi & lo limits alarm delay<12s |
Lead Off Condition |
Detected and displayed |
SpO2
| Display Range: | 0 to 100% |
| Accuracy: | 80-100%:±2digits; 7 0-69% Unspecified 0-80%:±3digits |
| Alarm Delay: | Spo2 hi & lo limits alarm delay<7s |
| Resolution: | 1% |
| Display Update: | <5s |
PULSE RATE
| Measurement Range: | 20 to 250bpm |
| Resolution: | 1bpm - Accuracy : ±1bpm |
| Alarm Delay: | Pulse rate hi & lo limits alarm delay<7s |
RR
Technique: |
Trans-thoracic impedance (RA-LL) |
Range: |
0-100 rpm - Resolution : 1 rpm |
Accuracy: |
±2 rpm (0-60rpm) |
Alarm Delay: |
Resp. rate hi & lo limits alarm delay<7s |
TEMP
| Technique: | Resistance |
| Channels: | One channel (2nd channel optional) |
| Scales: | F. Or C. |
| Range: | 0-50. - Resolution : 0.1. - Accuracy ±0.1. |
| Alarm Delay: | Body temperature hi & lo limits alarm delay<7s |
NIBP
Technique: |
Oscillometric method (with inflatable cuff) | ||||||||||||
Patient Types: |
Adult, Pediatric and Neonate | ||||||||||||
Cuff Inflation Time |
3-15 seconds depending on cuff size. | ||||||||||||
Cuff Inflation Pressure |
Initially 180mmHg for Adult/Pediatric (100 or 70 mmHg for Neonate). Subsequent inflation pressures determined by last measured systolic pressure. | ||||||||||||
Measurement Modes: |
Manual Immediate upon operator command AUTO Determinations automatically made with selectable intervals. STAT Determinations continues in 5 minutes | ||||||||||||
Measurement Range: |
|
||||||||||||
Pressure Resolution: |
1mmHg | ||||||||||||
Accuracy: |
|
||||||||||||
|
Standard Deviation: |
8mmHg | ||||||||||||
|
Determination Time: |
Typically 25seconds. Varies with patient's PR, pulse pressure and amount of artifact present. |
||||||||||||
Over Pressure Valve: |
Automatically releases cuff pressure if inflation Pressure exceeds 280mmHg (Adult) /150mmHg (Neonate). |
||||||||||||
Overtime Protection: |
Stop determinates if the measurement time exceeds 120s (Adult)/90s. (Neonate). | ||||||||||||
Alarm Delay: |
Pressure hi & lo limits alarm delay<7s |
POWER REQUIREMENTS
Power Supply |
100~240VAC,50/60Hz 1.6A max |
Internal Battery |
12V 7AH; Type- sealed lead-acid |
Operating Time |
2 hours typical at 25.,no printing, one NIBP |
(fully charged battery) |
measurement per 15 min. |
ENVIRONMENT
| Operating Temp: | 5. to 45 |
| Relative Humidity: | 0 to 80%, non-condensing |
| Height Of Sea Level | -500m ~ 4600m |
The transport and storage environment should comply with the following conditions:
| Storage Temperature: | -20. to 70 |
| Relative Humidity: | 0 to 93%, non-condensing |
| Height Of Sea Level: | -500m ~ 13100m |
OPTIONAL ACCESSORIES
Thermal Printer
IBP Modules
SPO2 Sensors
NIBP Cuffes
2nd Temperature Channel / Probe
For Further Details Contact:
Zeal Medical Pvt. Ltd.
4/19 -A, Piramal Indl. Est., S. V. Road, Goregaon (W),
Mumbai 400062. India
Tel:. 91 22 2876 02 74
TeleFax:. 91 22 2876 19 05
Email:. zealmedi@bom3.vsnl.net.in
Website:. http://www.zealmedical.com
Zeal Medical Pvt. Ltd.
4/19 -A, Piramal Indl. Est., S. V. Road, Goregaon (W),
Mumbai 400062. India
Tel:. 91 22 2876 02 74
TeleFax:. 91 22 2876 19 05
Email:. zealmedi@bom3.vsnl.net.in
Website:. http://www.zealmedical.com
Pediatric Oncall does not endorse any specific product, service, or treatment.
Partner Sites
Disclaimer:
The information given by www.pediatriconcall.com is provided by medical and paramedical & Health providers voluntarily for display & is meant only for informational purpose. The site does not guarantee the accuracy or authenticity of the information. Use of any information is solely at the user's own risk. The appearance of advertisement or product information in the various section in the website does not constitute an endorsement or approval by Pediatric Oncall of the quality or value of the said product or of claims made by its manufacturer.
copyright ©2011 website design & development by Levioza