Warmer with Fixed Baby Cradle (SKIN / MANUAL) & Under Surface PhototherapyMODEL: RHW1003B |
|
Micro Computer Temperature Controllers
- Provides thermal regulation with highest degree of accuracy
- User friendly feather touch control
- Safety alarms along with high temp. cutoff
SKIN / MANUAL (Micro Computer Temperature Controllers) |
|
1” bright skin temp. display
Set temp. display
% of heater output display
Skin mode / Manual mode |
| Manual Mode |
|
Time programmable
Heater output settable
Over temp. cutoff. |
| Timer / APGAR Timer Baby attending alarm Auto audio alarm mute Mute time programmable Safety Alarms Messages on LCD |
| Skin Mode |
|
|
Skin high 0.5° C
Skin low 0.5° C
Skin over range (>38° C) |
Skin under range (<34° C)
Skin probe failure |
| Air Mode |
|
Air high 1.5° C
Air low 3.0° C |
Air over range
Air probe failure |
|
Safety high temp. cutoff at |
|
Skin temp. >38° C
Air temp. >39° C |
| Heater & Power failure alarm |
Heater Box
- Consists of Ceramic Infra Red Heater.
- The source box can be swiveled to take X-ray
- Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
- The construction of the stainless steel reflector and guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards
Halogen Examination Lamp
- Provides accurate assessment of Infant colour and sufficient illumination during procedures
Baby Tray (Stainless Steel)
- Clear acrylic collapsible side panels to assess an infant from all sides
- Stainless steel baby tray with transparent slot to incorporate the facility of undersurface phototherapy
- Acrylic tray and cushion mattress with head up/down facility
- Acrylic X-ray tray with X-ray guide for positioning the cassette
- The design aspects also allow easy of dismantling baby care area during cleaning or fumigation
Essential Attachments
- Two Lower shelf, Side Mounted Monitor tray, Pole for Syringe Pump/ Infusion Pump & Stainless Steel IV Stand
|
|
|
|