Technical Specification
CONTROL PANEL
Display:
LED display of measurement results, instruction, troubleshooting messages, signal strength bar and time
Parameters Displayed: Systolic Pressure, Diastolic Pressure, Mean Arterial Pressure (MAP), Pulse Rate, % SpO2
PHYSICAL DIMENSIONS & WEIGHT
Base Unit : (15.0 cm H X 17.5 cm W X 8.0 cm D) Weight : 1.0 kg
NIBP
Technique: Oscillometric Application : Neonatal, pediatric and adult patients Blood Pressure Range : Systolic : 30-240 mmHg Diastolic : 15-180 mmHg MAP : 25-200 mmHg Pulse Rate Range : 30-250 bpm
Resolution :
NIBP : 1 mmHg Pulse rate : 1 bpm
OXIMETRY
ASpO2 : Anti-motion SpO2 SpO2 % Range 0-99 % SpO2 Accuracy : 70-99 % 2 digits 50-69 % 3 digits 50-99 % 3 digits (motion) Pulse Rate Range : 30-250 bpm
Pulse Rate Accuracy :
1 bpm 3 bpm (motion)
OPERATING MODES
History : Review of previous measurements NIBP : Manual or Automatic (at preset intervals) % SpO2 : Continuous monitoring
POWER
Source : External AC power or internal battery AC Power : 85-264 VAC, 50/60HZ, 0.5A 0.5A fuse rating Battery : Lithium Ion (rechargeable) Charge Time : 4 hours , Operating Time : 70 readings at 5 minute intervals Leakage Current : 100 " A (maximum)
PATIENT ALARMS
Adjustable high and low alarms for Pulse Rate, % SpO2, Systolic Pressure, Diastolic Pressure and mean Arterial Pressure.
SAFETY LIMITS
Automatic Cuff Deflation : If cuff pressure exceed 280 mmHg (Adult),
160 mmHg (Neonate)
OPERATING ENVIRONMENT
Operating Temperature : 0 to 500C
Storage Temperature : -20 to 650C
Humidity : 10-95%, non-condensing